MOISTURE BALANCE MOC12H
MOISTURE BALANCE MOC12H
SHIMADZU-UNIBLOC-ELECTRONIC-MOISTURE-BALANCE-MOC12H
Shimadzu-Unibloc-Electronic-Moisture-Balance-MOC12H supports Large sample pan to be placed for the best drying condition.
Shimadzu-Unibloc-Electronic-Moisture-Balance-MOC12H Reliable Unibloc Weighing mechanism and unique conditions auto-tare system assure accurate measurements.
Large Sample pan and continuous auto-tare mechanism
A large pan sample pan contribute to accurate measurements,but the larger heat capacity of oit normally results in a larger zero drift in the precious weighing mechanism which eliminates the zero drift continuously and assures high accuracy ,even with larger sample pan.
Unibloc technology for precision weighing
Shimadzu’s unibloc cell is used as the core mechanism of the weighing part.
its uniform structure maintains the high performance of precision weighing under repeated heating/cooling
Mid-wave infrared quartz heater
Min-wave infrared quartz provides effective drying for a wide range of sample.Besides the excellent drying performance ,it offeres a long operational life of 20,000 to 30,o00 hours.therefore the long term operation cost is much lower then halogen lamp heaters.
Predictive Measuring mode :
The Final result can be predicted from the drying process saving time in repeated from the drying process,saving time in repeated measurements.
WindowsDirect
Complete sample data and instrument setting can be directly typed into any application on windows and no interface software is required
Specification :
Model | MOC-120H |
Measuring method | Infrared heat drying mass measurement |
Sample pan size | ø130 mm |
Minimum display in weighing | 0.001g |
Measurement range of moisture content | 0.01% to 100.00% |
Moisture content minimum display | 0.01% |
Maximum sample capacity | 120 g |
Measurement mode | Automatic or Timed halting modes, Rapid, Slow and Step drying modes, Predictive Measuring mode |
Drying heater | Mid-wave infrared quartz heater |
Setting temperature range | 30°C to 180 °C (by 1°C increments) |
Dimensions & weight | W220xD415xH190 (mm), 4.5 kg |
Operational temperature and humidity range | range5 °Cto 40°C, 85% RH maximum |
Power requirements | 100 V to 127 V/220 V to 240 V AC, 640 W maximum |
Standard accessories | Sample pan 2 pcs, sample pan stop, windbreak case, sample pan, handler 2 pcs, aluminum sheet 20 pcs, spoon, spatula |